High Power DC Electronic Load 63200 Series Operation & Programming Manual
7.2.11 SPECification Sub-system
SPECification
:C
:L
:H
[:PASS]
:TEST
:UNIT
SPECification:C/L/H
Description Set the voltage (CC, CR, CP mode) or current (CV mode)
specification.
Syntax SPECification:H
SPECification:L
SPECification:C
Parameter For valid value range refer to respective specification.
Example SPEC:H <NRf+>[suffix]
SPEC:L <NRf+>[suffix]
SPEC:C <NRf+>[suffix]
Query Syntax SPECification:H?
SPECification:L?
SPECification:C?
Query Example SPEC:H?
Return Parameter <NR2> [CC,CR,CP mode Unit=Voltage], [CV mode Unit=Current]
Return Example 4.75
SPECification[:PASS]?
Description Request to refer all channels specifications for GO-NG result.
Query Syntax SPECification?
Query Example SPEC? Return all channels GO-NG result.
Return Parameter 0 | 1
Return Example 0 (NG), 1 (GO)
SPECification:TEST
Description Start or close the specification test.
Syntax SPECification:TEST <NR1 | CHAR>
Parameter 0 | OFF, 1 | ON
Example SPEC: TEST 1
SPEC: TEST OFF
Query Syntax SPECification:TEST?
Query Example SPEC:TEST?
Return Parameter 0 | 1
Return Example 1 (ON)
SPECification:UNIT
Description Set the specific entry mode.
Syntax SPECification:UNIT <NR1 | CHAR>
Parameter 0 | PERCENT, 1 | VALUE,
Example SPEC:UNIT 1
SPEC: UNIT PERCENT
Query Syntax SPECification:UNIT?
Query Example SPEC:UNIT?