Appendix E Specification Verification
The test result of each voltage range for 66200 Series Power Meter:
66200 Series 500V Range
Fluke 5500 Output Parameter Max. Spec. Measured Result Min. Spec.
High Voltage
500Vrms
60Hz
Vrms 500.9 499.1
Low Voltage
50Vrms
60Hz
Vrms 50.45 49.55
High Voltage
500Vrms
10K Hz
Vrms 503.4 496.6
66200 Series 300V Range
Fluke 5500 Output Parameter Max. Spec. Measured Result Min. Spec.
High Voltage
300Vrms
60Hz
Vrms 300.55 299.45
Low Voltage
30Vrms
60Hz
Vrms 30.27 29.73
High Voltage
300Vrms
10K Hz
Vrms 302.04 297.96
66200 Series 150V Range
Fluke 5500 Output Parameter Max. Spec. Measured Result Min. Spec.
High Voltage
150Vrms
60Hz
Vrms 150.27 149.73
Low Voltage
15Vrms
60Hz
Vrms 15.14 14.86
High Voltage
150Vrms
10K Hz
Vrms 151.02 148.98
Current Measurement Specification Verification
Steps:
1. Set the voltage range of 66200 Series Power Meter to 150Vrms and current range to 2A,
the current measurement display is set to I (AC current RMS).
2. Follow the test table to set the voltage, high current output of Fluke 5500A.
3. Press OPR to set the Fluke 5500A to begin output.
4. Log the current RMS showed on the 66200 Series Power Meter current measurement
display panel.