Digital Power Meter 66200 Series User’s Manual
5. Follow the test table to set the voltage, low current output of Fluke 5500A.
6. Repeat step 3 and 4 to log the 66200 Series Power Meter panel readings.
7. Set the Fluke 5500A output to Standby and follow the test table to switch the 66200 Series
Power Meter current range.
8. Repeat step 2 to 7 and measure the current of remaining ranges.
9. Once the test is completed, set the Fluke 5500A output to Standby.
a CAUTION
Make sure the Fluke 5500A output is OFF when switching the current range to avoid any
measurement error. Do not touch the test wire when performing Fluke 5500A output test to
prevent electric shock. Once the test is done, make sure the Fluke 5500A output is OFF
before changing the wire configuration to avoid electric shock.
The test result of each current range for 66200 Series Power Meter:
66200 Series 2A Range
Fluke 5500 Output Parameter Max. Spec. Measured Result Min. Spec.
High Current
100V/ 2Arms
60Hz
Irms 2.007 1.993
Low Current
100V/ 0.2Arms
60Hz
Irms 0.205 0.195
High Current
100V/ 0.2Arms
10K Hz
Irms 0.2062 0.1938
66200 Series 0.4A Range
Fluke 5500 Output Parameter Max. Spec. Measured Result Min. Spec.
High Current
100V/ 0.4Arms
60Hz
Irms 0.4014 0.3986
Low Current
100V/ 0.04Arms
60Hz
Irms 0.0410 0.0390
High Current
100V/ 0.04Arms
10K Hz
Irms 0.40124 0.39876
E-4