- 18 -
Reaction of the balance:
- Activity of the balance (eg. the balance calibration) is done for submenu described
as Function
- Attribute activation for submenu which is indicated (digit flashing means the value
can be changed and some signs can be written)
Draw. 8. The balance submenu
7. WEIGHING
Follo
wing conditions must be fulfilled to get reliable results:
- Stable temperature
- Stable ground
- Proper parameters for external conditions
Before measurements or for essential changes of the external conditions (if the
temperature changes more than 1° C/h) calibrate the balance in accordance with p.
7.1.
Before measurements load the pan and check if the balance show „precise zero” –
displayed in down left corner of the display (only if the parameter P4 06
Autozero has the value 1: yes) and check if the measurement is stable – is
displayed in right up corner of the display. If the balance does not show zero press
the key
If the conditions are unfavourable (no stable result) lines appear on the display.
After settled time the balance returns to weighing mode without set up to zero. In
this case wait untill the conditions stabilize and press Esc again
By the Units key select measure unit. Put the load on the pan and after stabilization
read out the result. If measure unit user wants to use is not displayed during
pressing the Units key check if it has access attribute.