- 28 -
Apart from information settled in menu group the report contains: calibration mass
remembered by balance after last calibration (description Old:), calibration mass
determined during actual calibration (description: Calibration) and deviation of the
calibration (description Deviation:) – difference between these two masses.
Draw. 17. Example of balance calibration report
9. SETTING PRINTOUTS FOR GLP PROCEDURES
P2 GLP is grou
p of the parameters which declares factors on the calibration printout. For
fields:
- user (max 8 alphanumerical signs)
- design (max 8 alphanumerical signs)
introduce names by the balance keyboard or the PS/2 keyboard. For the rest select:
- 1 no (do not print during report)
- 0 yes (print during report)
Main view of the GLP submenu is presented on drawing 19, page 24. If you use the PS/2
keyboard see what dependences are between the balance keyboard and PS/2 keyboard
(p. 5.1.2)
10. SETTING TIME AND DATE
T
he balance has real time clock whose parameters can be modificated. Enter the menu
group P3 Date/Time as it is show below: