- 56 -
17.8. Statistics
Activation
Draw. 52. Statistics –function activation
Results of previous statistics should removed after function activation. It is realized through
option M8 01 Cancel.
All statistic date are actualized after write following measurement to the balance memory.
Following measurement is written to series after load is put on the pan, stabilization of the
result (measure unit is displayed) and after pressing ENTER.
User decides what statistic date are presented on the graphic display during
measurements by setting their activity in the submenu of work mode (values which are set
for YES are active). Independly on setting (YES/ NO), during final result the printout
contains full statistics.
N : 5 (quantity of weights)
SUM : 161.121 g (all components total mass)
X : 32.224 g (average mass of weighed components)
MIN : 20.486 g (min mass)
MAX : 35.578 g (max mass)
D : 15.092 g (difference between Max- Min)
SDV : 6.581 g (standard deviation)
RDV : 20.4 % (variation factor)