- 26 -
8.3.2. External calibration
T
he external calibration should be performed with external mass class:
- F2 – for balances CX
List of weights for separate balances is included in technical specification in the final part of
the manual.
1. Move to submenu P1 – Calibration.
2. Select the function 02 external calibration
3. Press the RIGHT ARROW key.
4. Order to take the load off the pan appears on the display (no load on the pan).
After yesing load off the pan press the ENTER key.
5. The balance determines mass of empty pan
6. Put load and press the ENTER
7. After the calibration the balance returns to submenu P1 - Calibration
8. Return to weighing – as in the point 6.1.2.
If the function DRH is active user cannot perform the external calibration
process. Function DRH is active in verified balances.
8.3.3. Calibration performed by user
User cali
bration should be made with any external weight of class:
• F2 – for balances CX
- Enter menu group P1 Calibration. Select the parameter 03 user calibration
- Press the RIGHT ARROW key. The balance displays order to note calibration mass.
The first digit flashes and it can be changed.
Draw. 14. User calibration – declaring value of weight
- Record new external mass by functional keys (in accordance with p. 6.1.1 of the
manual)
- Confirm the mass. The balance starts calibration and shows orders on the display.
- The balance determines mass of empty pan and shows order to put this mass