Programmable Voltage Range 0.10 kV-6.00 kV (10V/Step )
Voltage Display and Accuracy 0.10 kV-6.00kV ±1% ± 5 Counts
Voltage Display Resolution 0.01kV
Current Display R and Accuracy 0.01mA-10.00mA ±1% ± 5 Counts
Current Display Resolution 0.01mA
Selectable Range of Pass/Fail Levels* 0.01mA-10.00mA
Maximum Current Output 10.00mA
Optional Arc Detection 9 Levels
* Full selectable range only achievable with the use of Safety Base, excluding HAL10
2
excluding HAL102
Programmable Voltage Range 0.10 kV-5.00 kV (10V/Step)
Frequency Independent of Supply 50 or 60 Hz
Voltage Display and Accuracy 0.10 kV-5.00kV ±1% ± 5 Counts
Voltage Display Resolution 0.01kV
Current Display and Accuracy 0.01mA-20.00mA ±1% ± 5 Counts
Current Display Resolution 0.01mA
Selectable Range of Pass/Fail Levels* 0.01mA-20.00mA
Maximum Current Output 20.00mA @5kV
Optional Arc Detection 9 Levels