Single Phase Power Measurement
Display Range and Accuracy 0.02kVA-5.00kVA ±2% ±0.02kVA
Selectable Range of Pass/Fail Levels 0.001kVA-5.000kVA
Power Output Rating
Test Voltage 110V – 230V AC 20A nominal (Vin-
Vout)
Maximum Power Output
Up to 5.0kVA (dependant on mains
supply rating)
Single Phase Power Factor Measurement
Display Range and Accuracy Ratio 0.000 -1.000 ± 0.030
Selectable Range of Pass/Fail Levels 0.000-1.000
Leakage Current Measurement
Display Range and Accuracy 0.10mA-20.00 mA ± 1% ±5 Counts
Display Resolution 0.01mA
Selectable Range of Pass/Fail Levels 0.01mA-20mA
Touch Current Measurement
Display Range and Accuracy 0.02mA - 5.00 mA ± 1% ±5 Counts
Display Resolution 0.01
Selectable Range of Pass/Fail Levels 0.02mA-5.00mA
Measuring Device
IEC/EN 60990 Fig 4
(IEC/EN 60990 Fig 3 or Fig 5 Upon Request)