106
1 Take a fresh Super Sani-Cloth
®
Germicidal
Disposable Wipe from the container.
Remove excess moisture from the wipe.
With the meter turned off, wipe the meter’s
exterior clean.
Do not let liquid accumulate near any open-
ing. Carefully wipe around the cover of the
test strip guide and other openings. Make
sure that no liquid enters the meter.
To properly disinfect, let the meter sit wet
for a contact time of at least 2 minutes,
or as instructed by the manufacturer in
the product labeling. Always refer to the
manufacturer’s labeling for contact time
recommendations.
2 With a fresh dry cloth or lint-free tissue,
wipe away residual moisture and uids
after cleaning the housing.
3 Allow wiped areas to dry for at least 10
minutes before performing a test.
Cleaning and Disinfecting the Meter
CoaguChek INRange User’s Manual