110
On completion of cleaning:
4 With the cover off, let the test strip guide
dry for at least 10 minutes before re-
attaching the test strip guide cover and
testing again.
Visually verify that no residual moisture is
seen anywhere on the test strip guide and
cover at the completion of cleaning.
Ensure that the test strip guide and cover
are completely dry before assembling
them.
WARNING
5 Re-attach the test strip guide cover to the
housing. Make sure that the cover is prop-
erly closed. You will hear it snap into place.
Cleaning and Disinfecting the Meter
CoaguChek INRange User’s Manual