Ideally the flue/ duct should be operating under clean conditions (zero opacity)
for commissioning. If this is not practical, however, a target calibration opacity
can be assumed, in which case, should a clean flue become available, the
instrument must be re-calibrated.
! Before completing commissioning and leaving equipment
operational, ensure that the air supply to the purges is
operating.
Allow enough time for each key press to action otherwise double keystrokes
may be entered.
Initial Calibration
• Press MODE 4 times, to access SET UP mode.
• Press ENTER 4 times, to enter 0000 for the security code.
• Press DOWN arrow 8 times and ENTER once, to access
CALIBRATE option.
• Press DOWN arrow twice and ENTER once to access
CALIBRATE.
• Set opacity Target to zero (for clean duct) or assumed opacity. Do
this by using the UP and DOWN arrow keys.
• Press ENTER, the menu will now enter the Cycles option. Set this
to the required number (30 is recommended for this initial
calibration).
• Press ENTER twice and the calibration will start, the calibration
factor will be given at the end of the process and the instrument
will return to normal operating mode.
Operating Parameters
For the instrument to function as required, all operating parameters must be
entered into the microprocessor within the signal processor. The operating
parameters of the instrument are set from the set-up mode, they are then held
in non-volatile memory and retained in the event of a power loss.
! All operating parameters are entered in the set-up mode - see
the ‘Operation’ section for details.