Identification
From this option, the analyser type, its program number, and the device identity
of the instrument may be displayed.
• Instrument type - EnergyTech 100 Series 101/102
• EPROM number - will display the communications EPROM ID
• Identity - for factory set-up only
Parameters
The following parameters are examined from this option, selected using the
ARROW keys :
• Measurement Path Length
The path length between the transmitter and receiver. This value is not used
within the current program and is for reference only.
• Dust Factor
A 'Dust Factor' is used to calculate the dust density; see section 5.12.
• Output Fault 1
Should a fault condition occur on output 1, the analogue output can be set from
one of four options see the set-up mode. The selected option may be examined
here.
• Output Fault 2
As above for output 2.
• Auto Zero
The auto zero facility compensates for any optical contamination.
Compensation ranges from 0.1% - 8% and can be displayed here. INHIBIT is
displayed if the option is not selected.