CODEL
OPS. 141 20
Issue : A
Revision : 1
Date : 14/06/18
Doc i/d : 0141/6
050028
! Make a note of this number; this is the security code required
every time the SET UP MODE is entered in the future.
31.Press the DOWN arrow key to select Identity, press ENTER. Use
the arrow keys to scroll the device identity number between 01 to
255.
32.Press ENTER and the DOWN arrow to select Pathlength. Press
ENTER.
33.As before alter each digit individually using the arrow keys; press
ENTER when the correct value is selected. After the fourth digit
the menu will return to Pathlength.
34.Use the DOWN arrow to select Dust Factor sub menu.
35.Press ENTER to access the option and use the arrow and ENTER
keys as usual to select the value required.
36.Press ENTER when complete.
37.Use the DOWN arrow to access Auto Zero option. Set the figures
to the required value.
38. Press the DOWN arrow and the EXIT option is accessed on the
LCD. Press ENTER to leave the PARAMETERS menu. Use the
DOWN arrow key to select the NORMALISATION menu.
39. Use the same key combinations to select all normalising
parameters, if required. EXIT this menu to return to SET UP mode.
40. Use the DOWN arrow key to access the CALIBRATE menu.
Press ENTER.
41. Use the DOWN arrow to select the CALIBRATE option. Proceed
as described previously. See section 3.6. Basic Calibration.
42. Note the new Cal Factor and detector levels.
43. The unit is now fully commissioned.