Kit Installation
2-2 February 2016 | 256574-ENG
R01
2.1.1 Unpacking
1. Remove the wooden crate by removing the screws, starting at the top of the crate.
2. Using a utility knife, carefully remove the cardboard packing around the chamber.
3. Check the shock indicator on the back of the chamber for indications of damage during
shipping. Contact Conviron immediately if the shock indicator is red.
4. Remove the chamber from the shipping pallet following the instructions included on the
pallet.
5. Verify receipt of all components per the packing list. Notify your Dealer or Conviron
Sales member immediately of any missing components.
6. Ensure all doors are locked before moving the chamber to the installation location.
Keep the caps on the unused lamp receptacles at all times.
A1000 chambers can operate at high temperature and humidity creating an
electrocution hazard if the receptacle is exposed.
When replacing kits, ensure all unused receptacles are capped when not in use.
The finish and color of the parts received with this kit may not exactly reflect the parts
shown in these instructions, due to product updates. However, the installation
instructions are accurate.