Maintenance
7-2 February 2016 | 256574-ENG R01
7.3 Replacing Lamps
Periodically check to ensure all lamps functioning properly. Replace poorly lit or flickering lamps to
ensure optimal unit performance.
To replace a fluorescent lamp:
1. Identify the lamps to be replaced.
2. Completely disconnect the power supply to the unit by unplugging it from its receptacle
before replacing lamps.
3. Unlock the lamp to be replaced by rotating it ¼ turn in its socket and remove with care
to avoid breaking it.
4. Install the new lamp.
5. Ensure it is locked in its socket by rotating it ¼ turn.
6. Dispose unused lamps, following the requirements in your area or contact your local
authorities for procedures.
7.4 Ultrasonic Humidity System
When not in use, the humidity box (located inside the machine compartment), must be drained
and cleaned. Close the water supply valve while not in use.
Conviron recommends setting a moderate RH value in all programs to keep the USH feature
operable without significantly affecting the experiment.