Instruction Manual | 29
Warning/Information Message Meaning/Recommendations
error 40 - Camera power supply control failure. Try to restart the Videodrop and qvirCLS. If the error appears
again, contact Corning support team (Videodrop@corning.com)
or your local representative.
error 41 - Communication failure with camera.
Please click on Record to restart measurement.
Several images have been lost during the data transfers from the
camera. Click on Record to restart measurement.
If this error occurs often, verify USB camera connection cable
and/or contact Corning support team (Videodrop@corning.com)
or your local representative.
Should you have a question about the operation of the Corning® Videodrop, qvirCLS, or if service is required, please email
Videodrop@corning.com.
8.0 Switch Off the Device
Always switch Off the device in the following sequence:
Remove the sample holder from the Videodrop, and discard it accordingly with adapted waste disposal procedure.
Position the LED protection cap on the LED before positioning the handle in upper position.
Exit qvirCLS software.
Switch Off the Videodrop.
Turn Off the computer.
Position the slipcover on the Videodrop.
Refer to the warning in Section 2 regarding turning Off the device.
9.0 Cleaning
Periodic cleaning and disinfection of the Videodrop is recommended depending on its usage and environment.
9.1 General recommendations
Always use eye protection and gloves to avoid injury. Respect the use-by date of cleaning products and disinfection solutions.
Ensure that the contact time and concentration of cleaning product and disinfection solution are appropriate for the equipment
and the samples being used. Carefully observe the instructions provided on the label of the cleaning product and the disinfection
solution.
Read the Association for Professionals in Infection Control and Epidemiology (APIC) and Food and Drugs Administration (FDA)
recommendations carefully if applicable in your country.
9.2 Cleaning the device
When cleaning, switch Off, and disconnect the device to prevent electric shock.
Do not spray cleaning or disinfectant directly on the device. Leaks could damage the system.
Do not use liquid on electronic parts (LED, connectors). Do not let liquid enter inside the device.
Surfaces must be cleaned in strict compliance with the following steps:
Clean using a soft cloth soaked in the recommended cleaning product.
If necessary, rinse using a soft cloth soaked in water.
Wipe the surface with a soft cloth soaked in the recommended disinfection solution.
If necessary, dry thoroughly with a soft, clean and absorbent cloth.
If damage is found, stop using the device, and contact Corning.