68
Advanced operations
Modify the order of sending in the Memory Map.
Number the instrument in the PROFIBUS net “Node number”.
Initialize the communication.
5.3.5 Channel status
5.3.6 Measuring units: 1 byte
When the bit with the biggest weight (0x80) is activated, it indicates sensor alarm.
Table 5-1Channel status
Bit Value Description
0 0x1 Limit 1 active
1 0x2 Limit 2 active
2 0x4 Alarm active
3 0x8 Measurement error (for example, measurement out of range)
4 0x10
Temperature error (measurement out of range or probe not
connected)
5 0x20
Cleaning relay active (while cleaning the measurements are
blocked)
6 0x40
Intermittent bit to detect active communication of the
instrument (must change at least 1 time each 2 seconds)
7 0x80
Channel bit active. Indicates that the measuring board is
connected and the channel is active. Calibrating or
programming the channel, this bit is 0
Table 5-1Measuring units: 1 byte
Value Description
0pH
1 mV (Channel pH/mV)
2 Conductivity, S/cm.
3TDS, g/L
4 NaCl, g/L
5 Resistivity, Ohms x cm
6 Dissolved oxygen, ppm
7 Dissolved oxygen,%
8 Dissolved oxygen, mg/L
9 Chlorine, mg/L
10 Chlorine, ppm
11 Chlorine, mV
12 °C (Channel pH/mV measuring and controlling °C)
127 No connected electronic board