326Appendix F: Technical Information
No cleaning methods are recommended or tested for the receiver. The useful life
of the receiver is 1 year. The service/useful life for the peripherals is 1 year. If you
have difficulty reading your receiver in bright sunlight, you may need to seek a shady
location. Do not connect the receiver to any equipment not specified in IFU.
Guidance and Manufacturer’s Declaration –
Electromagnetic Immunity
The receiver is intended for use in the electromagnetic environment specified in the
next table. The customer or the user of the receiver should ensure that it is used in
such an environment.
Receiver Electromagnetic Immunity Specifications
Immunity Test
IEC 60601 Test
Level
Receiver
Compliance Level
Electromagnetic
Environment
Guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
± 8 kV Contact
± 15 kV Air
± 8 kV Contact
± 15 kV Air
Floors should be
wood, concrete or
ceramic tile. If floors
are covered with
synthetic material,
the relative humidity
should be at least
30%.
Electrical Fast
Transient/Burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for
input/output lines
± 2 kV for power
supply lines
Not applicable
Main power quality
should be that of a
typical commercial or
hospital environment.
Surge
IEC 61000-4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
± 1 kV line(s) to
line(s)
Not applicable
Main power quality
should be that of a
typical commercial or
hospital environment.