327Appendix F: Technical Information
Receiver Electromagnetic Immunity Specifications
Immunity Test
IEC 60601 Test
Level
Receiver
Compliance Level
Electromagnetic
Environment
Guidance
Surge
IEC 61000-4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
± 1 kV line(s) to
line(s)
Not applicable
Main power quality
should be that of a
typical commercial or
hospital environment.
Voltage
Dips, Short
Interruptions
and Voltage
Variations on
Power Supply
Input Lines
IEC 61000-4-11
IEC 60601-1-11
0% U
T
for 1 cycle
0% U
T
for 0.5 cycle
at 8phase angles
70% U
T
(30% dip in
U
T
) for 25 cycles
0% U
T
for 250 cycles
0% U
T
for 1 cycle
0% U
T
for 0.5 cycle
at 8phase angles
70% U
T
(30% dip
in U
T
) for 25 cycles
0% U
T
for 250
cycles
Main power quality
should be that of a
typical commercial or
hospital environment.
Power
Frequency
(50/60 Hz)
Magnetic Field
IEC 61000-4-8
30 A/m 30 A/m
Magnetic fields from
common appliances
are not expected to
affect the system.
NOTE: U
T
is the AC main voltage prior to application of the test level.
Guidance and Manufacturer’s Declaration –
Electromagnetic Immunity
The G6 is intended for use in the electromagnetic environment specified in the next
table. The customer or the user of the G6 should ensure that it is used in such an
environment.