Dexcom G6
®
System User Guide
328Appendix F: Technical Information
System Electromagnetic Immunity Specifications
Immunity Test
IEC 60601 Test
Level
System
Compliance
Level
Electromagnetic
Environment Guidance
Conducted RF
IEC 61000-4-6
(Receiver only)
3 Vrms
150 kHz to 80
MHz
6 Vrms
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Recommended Separation
Distance
d = 1.2 √P 150 kHz to 80 MHz
d = 1.2 √P 80 MHz to 800 MHz
d = 2.3 √P 800 MHz to 2.5 GHz
Where P is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and d
is the recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined
by an electromagnetic site
survey,
a
should be less than
the compliance level in each
frequency range
b
.
Interference may occur in the
vicinity of equipment marked with
the following symbol:
Radiated RF
IEC 61000-4-3
10 V/m at 80
MHz to 2700
MHz (AM
Modulation)
10 V/m