Operating Instructions RF-2000
15
4 Installation
4.1 Basic Installation Requirements
Warning:
To take full advantage of the RF-2000 performance capabilities and
to ensure its operational stability over a long service life, verify that
the selected installation site satisfies the following requirements.
1. Ventilation Ventilate the room where the high performance liquid
chromatograph is located since the solvent used is
flammable and/or toxic.
2. Fire Never use fire in the same room where the high
performance liquid chromatograph is installed. Also, avoid
installation in the same room of other devices which may
spark. Always keep a fire extinguisher nearby in case of
accident.
3. Sink Install a sink nearby for flushing eyes or skin which have
been in contact with solvent.
4. Corrosive gas and
dust
Avoid installation in a place exposed to corrosive gases or
dust.
5. Electromagnetic noise Avoid locations subject to intense magnetic or
electromagnetic fields. Use an additional noise filter if
power line noise interferes.
6. Space requirements This system is designed to be used on a table or stand,
preferably a solid and flat surface with a depth of 70cm or
more.
7. Others Select an installation site with the following parameters to
maintain full performance of the system.
(1) Maintain room temperature within 4~35C, without
extreme fluctuations.
(2) Avoid direct output of a heater or a cooler.
(3) Avoid exposure to direct sunlight.
(4) Avoid locations subject to strong vibrations or
prolonged weak vibrations.
(5) Maintain relative humidity within 45~85%.
(However, humidity should be below 70% when the
room temperature is 30C or higher.)