Operating Instructions RF-2000
80
9.2 List of Error Messages
9.2.1 Errors which cannot be reset
Turn the power of the instrument OFF and turn it ON again. If the error message
persists, turn OFF the power and contact your Dionex Service Representative.
(1) RAM error
It is displayed when something is wrong with the RAM.
EX(nm) EM(nm) Data
RAM FAILURE
(2) ROM error
It is displayed when something is wrong with the ROM.
EX(nm) EM(nm) Data
ROM FAILURE
(3) OVERHEAT error
It is displayed when the temperature in the instrument is abnormally high. If
this error occurred, the Xenon lamp is not lit even if the power is turned ON
again.
To repair this instrument and to cancel this error, contact our sales office or
representatives.
9.2.2 Errors which can be reset
Alarm sound stops by pressing CE and the next step is proceeded.
(1) Set value loss error
It is displayed if the parameters or the time program set previously are lost at
power on. Press CE and the instrument becomes operational. Set parameters
again.
EX(nm) EM(nm) Data
NOT PROTECTED
(2) File type error
It is displayed when the spectrum type or wavelength range for the file 0 and
those for the file 1 or 2 which is to be output in SPC PLOT are different.
EX(nm) EM(nm) Data
ERR FILE TYPE