5.5 Additional Functions
AUX. FUNC (AUXILIARY FUNCTION) is equipped in this instrument.
5.5.1 AUX. FUNC List
Type Command Function
4 TIME Displays the elapsed time while a time program is
executed (when MONIT-TIME is set)
4 SMPL EN Displays output from emission measurement sensor
4 REF EN Displays excitation light beam monitor value
1 GAIN Sets gain. (1 = x1, 2 = x4, 3 = x16)
1 SCAN FILE Sets file No. Where scan data is to be stored
(0=Background, 1 and 2=Sample)
1 SPC TYPE Sets scan type. (1=EX, 2=EM)
1 EX SCAN BGN Sets excitation scan start wavelength. (200 ~ 900nm)
1 EX SCAN END Sets excitation scan end wavelength. (200 ~ 900nm)
1 EM SCAN BGN Sets emission scan start wavelength. (200 ~ 900nm)
1 EM SCAN END Sets emission scan end wavelength. (200 ~ 900nm)
1 SCAN SPD Sets scanning speed. (1=SUPER, 2=FAST,
3=MED,4=SLOW)
1 PLOT SPD Sets output speed of spectrum data to recorder.
(1=1nm/sec, 2=5nm/sec, 3=10nm/sec)
3 SPC PLOT Outputs spectrum data to recorder.
1 EVENT Sets EVENT OUT terminal status. (0=OFF, 1=ON)
1 EXT-S Used when EVENT OUT terminal is used for starting
external equipment, etc.
(0=EVENT, 1=Time program, 2=Unreleasable error,
3=Wavelength scan)
1 MONIT-TIME Sets when elapsed time is desired to be displayed in
executing a time program. (0=not to display, 1=to display)
1 LAMP Sets ON/OFF of Xe lamp. (0=OFF, 1=ON)
1 RATIO Decides operation of light source compensation by
division.
(0= without division, 1 = with division, 2 = Excitation light
energy)
4 XE TIME Displays accumulated operating time of xenon lamp.
1 Z WAVE Sets the output holding function when setting of
wavelength, sensitivity and gain is changed. (0=set, 1=not
set)
3 RS232C Sets communication parameters of RS-232C and activates
remote control mode. Refer to "7.2 Control from external
system via RS-232C".
3 CLOSE KEY Refuses any key input.
3 LAMP ADJUST Enters lamp position adjusting mode.
3 S/N CHECK Used for Performance Check.
"Type" in the above table indicates a kind of operating procedure.
Type 1: Enter a desired value with numerical keys and press the Enter key.
Type 3: Press the Enter key to execute the function.
Type 4: The status is displayed without any key operation.