6 Performance Verification
Inspection Environment
Carrying out inspection measurements in an environment where the equipment is
exposed to sudden changes of temperature or to air currents may affect the
repeatability of the data.
Inspections should be conducted in locations where changes of room temperature
are less than 2°C. Places where the equipment is exposed to strong air currents from
air conditioners should particularly be avoided.
Appropriate Location For proper and safe use of the equipment, select a site
in consideration of the following.
1. Corrosive Gas and Dust To extend the life of your equipment and maintain the
proper performance standards, avoid using your
equipment at a site exposed to corrosive gas or dust.
2. Ventilation and Fire If flammable liquids or toxic solvents are used, be sure
to ventilate the room well. Never use fire in the room
when flammable solvents are used.
3. Electromagnetic Noise Never use your equipment near devices that generate
strong magnetism. If the electrical power line generates
electrical noise, connect your equipment to a power line
with a commercially available filter.
4. Workspace Your equipment is designed to be used on a desk or
similar surface. Use only a firmly fixed, flat surface that
is at least 70 cm deep. If another piece of equipment is
to be placed to the left of the unit, be sure to provide at
least 15mm of space between the units.
In order to maintain the performance of your equipment, avoid subjecting it to:
(1) Direct air currents from air conditioners
(2) Direct sunlight
(3) Vibration
(4) High temperature or humidity. The room temperature should be within the
range of 4°C to 35°C, and the humidity between 45% and 85%, with minimum
fluctuation in room temperature. (However, humidity should be below 70%
when the room temperature is 30°C or higher.)