Operating Instructions RF-2000
36
5.4 Creating and Executing a Time Program
5.4.1 Command list
Commands which can be used for a time program are listed below:
Command Description Setting range Remarks
EXλ
Excitation wavelength 0, 200 ~ 900
EMλ
Emission wavelength 0, 200 ~ 900
ZERO Execution of zero
adjustment
None
MARK Marking on output for
recorder
None
RANG Output range for recorder 0 ~ 9 Recorder is short-
circuited when it is set
to 0.
RESP Response 1 ~ 4
SENS Sensitivity 1 HIGH
2 MED
3 LOW
GAIN Gain 1 x 1
2 x 4
3 x 16
SCAN Execution of wavelength
scanning
0 ~ 2 Specifies file no. where
scanned data is to be
stored.
EVNT EVENT output of ON/OFF 0 OFF
1 ON
LOOP Repeats time program from
the beginning.
0 ~ 255
0 means 256
times.
STOP Stops time program. None
LAMP Turn ON and OFF the
lamp.
0 OFF
1 ON
Example of time program is shown below:
EX(nm) EM(nm) Data
0.01 ZERO
Emission intensity is set at 0 in 0.01 minute after starting the time program.
EX(nm) EM(nm) Data
5.00 EX λ
λλ
λ 300
In 5minutes, EX (Excitation wavelength) is set at 300nm.
EX(nm) EM(nm) Data
5.00 EM λ
λλ
λ 400
In 5 minutes, EM (Emission wavelength) is set at 400nm.
EX(nm) EM(nm) Data
5.10 SENS 3