EasyManua.ls Logo

Dionex RF-2000 - Page 45

Dionex RF-2000
88 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operating Instructions RF-2000
43
5.5.2 Setting procedure for AUX. FUNC
TIME
(Display of elapsed time of time program)
EX(nm) EM(nm) Data
TIME
100.00
This function displays the elapsed time from the program start while the time program
is executed.
However, this is displayed only when MONIT-TIME is set to 1 in AUX. FUNC.
SMPL EN
(Output from emission measurement sensor)
EX(nm) EM(nm) Data
SMPL EN XXX.X
This function displays output from emission measurement sensor (Photo multiplier
tube).
REF EN
(Excitation light beam monitor value)
EX(nm) EM(nm) Data
REF EN XXX.X
This function displays output from excitation light beam monitor (Photodiode).
GAIN
(Setting of gain)
EX(nm) EM(nm) Data
GAIN __
This function sets gain for signal from the detector. Enter a desired value with
numeric key and press Enter.
Default set value at the factory is "2" (Gain = x 4). For general use, "2" (Gain = x 4)
can be applied.
Set value Gain
1 x1
2 x4
3 x16