Operating Instructions RF-2000
44
Magnification can be set as follows by combining with SENS (refer to section "5.2.5
Setting Sensitivity").
SENS Sensitivity
magnifications
3 (LOW) approx. x 1
3 (LOW) approx. x 4
3 (LOW) approx. x 16
2 (MED) approx. x 32
2 (MED) approx. x 128
2 (MED) approx. x 512
1 (HIGH) approx. x 1024
1 (HIGH) approx. x 4096
1 (HIGH) approx. x 16384
Please note:
When values of sensitivity (SENS) and/or Gain (GAIN) are changed,
it is necessary to create a new calibration curve.
SCAN FILE
(Setting of scan file)
EX(nm) EM(nm) Data
SCAN FILE __
This function sets No. of file where scan data is to be stored. Enter the desired file
No. with numerical key and press Enter.
Set value Type
0 Background
1, 2 Sample
SPC TYPE
(Setting of scan type)
EX(nm) EM(nm) Data
SPC TYPE __
This function sets spectrum scan type. Input a set value with numeric key and press
Enter.
Set value Type
1 Excitation scanning
2 Emission scanning
EX SCAN BGN
(Setting of excitation scan start wavelength)
EX(nm) EM(nm) Data
EX SCAN BGN _ _ _
This function sets start wavelength for excitation scanning.
Enter the desired start wavelength with a numeric key and press Enter (200 ~ 900).