Operating Instructions RF-2000
45
EX SCAN END
(Setting of excitation scan end wavelength)
EX(nm) EM(nm) Data
EX SCAN END _ _ _
This function sets end wavelength for excitation scanning.
Enter the desired end wavelength with a numeric key and press Enter (200 ~ 900).
Please note:
EX SCAN BGN should be smaller than EX SCAN END.
EM SCAN BGN
(Setting of emission scan start wavelength)
EX(nm) EM(nm) Data
EM SCAN BGN _ _ _
This function sets start wavelength for emission scanning.
Enter desired start wavelength with numeric key and press Enter (200 ~ 900).
EM SCAN END
(Setting of emission scan end wavelength)
This function sets end wavelength for emission scanning.
EX(nm) EM(nm) Data
EM SCAN END _ _ _
Enter desired end wavelength with numeric key and press Enter (200 ~ 900).
Please note:
EM SCAN BGN should be smaller than EM SCAN END.
SCAN SPD
(Setting of scanning speed)
EX(nm) EM(nm) Data
SCAN SPD __
This function sets scanning speed.
Enter the desired set value with numeric key and press Enter.
Set value Scanning speed
1 SUPER (approx. 3000nm/min)
2 FAST (approx. 600nm/min)
3 MEDIUM (approx. 120nm/min)
4 SLOW (approx. 24nm/min)
Please note:
When scanning speed is set to super, spectrum data is stored
at an interval of 4nm. When scanning speed is set to the other
value, spectrum data is stored at an interval of 1nm.
PLOT SPD
(Setting of plot speed)
EX(nm) EM(nm) Data
PLOT SPD __