Operating Instructions RF-2000
81
(3) No file error
It is displayed when there is no data in the file in SPC PLOT or there is no
data in the file 0 when the file 1 or 2 is specified (subtracted spectrum output).
EX(nm) EM(nm) Data
DATA NOT EXIST
(4) Peak detection error
It is displayed when the Raman peak of the distilled water is not detected for
S/N CHECK.
See item 6 of the table of "9 Troubleshooting" to solve this problem.
EX(nm) EM(nm) Data
NO PEAKS
(5) EEPROM error
It is displayed when something is wrong with the EE_PROM.
EX(nm) EM(nm) Data
EE-PROM FAILURE
(6) Home position error
It is displayed when the motor home position sensor does not operate
properly.
EX(nm) EM(nm) Data
ERR EX HOME POS1
Excitation motor origin (coarse)
EX(nm) EM(nm) Data
ERR EX HOME POS2
Excitation motor origin (fine)
EX(nm) EM(nm) Data
ERR EM HOME POS1
Emission motor origin (coarse)
EX(nm) EM(nm) Data
ERR EM HOME POS2
Emission motor origin (fine)