Operating Instructions RF-2000
82
(7) Wavelength calibration data error
It is displayed when the wavelength calibration data entered in the unit is
abnormal.
EX(nm) EM(nm) Data
ERR EX OFFSET
Excitation side
EX(nm) EM(nm) Data
ERR EX W TABLE
Excitation side
EX(nm) EM(nm) Data
ERR EM OFFSET
Fluorescence side
EX(nm) EM(nm) Data
ERR EM W TABLE
Fluorescence side
(8) LAMP error
It is displayed when the xenon lamp is not lit.
To lit the xenon Lamp again, operate the following procedure. Turn the
power OFF and then turn the power ON.
Set the "LAMP" parameter in additional functions to "1".
If this error message is displayed again, refer to "8.2 Checking the Xenon
Lamp" and "9 Troubleshooting".
EX(nm) EM(nm) Data
ERR LAMP
(9) LEAK error
It is displayed when leak of liquid from the flow cell is detected. See "8.1
Maintenance of the Flow Cell".
EX(nm) EM(nm) Data
ERR LEAK
Please note:
If the detector of the leak sensor is wet, even though there is
no liquid leakage, LEAK error may be indicated.
To recover this, refer to "8.1.5 Countermeasure to LEAK error".