Operating Instructions RF-2000
4
Important:
Be sure to read this instruction manual before supplying power to this unit.
• Be sure to mount the xenon lamp correctly before supplying the power to this unit
(refer to 4.7 Installing the Xenon Lamp). Failing to do so may damage the main
unit because high voltage around 30kV is applied to electrode when turning ON
the xenon lamp.
• Use the power supply voltage at the state which conforms to the power supply
setting (display of the power supply connector unit at the rear) of this unit. For
changing the power supply setting of this unit, refer to 8.4 Changing the Set Power
Supply Voltage. However, when the voltage is 220V ± 10%, set the power supply
voltage of this unit at 230Vac.