EasyManua.ls Logo

Distek Evolution 6300 - Calibrate Using Tempchek Digital Thermometer; Calibrate Using NIST-Traceable Reference Thermometer

Distek Evolution 6300
116 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 3: Operation
Operation Manual: Evolution Series
3
-
19
Calibrate Using TempChek Digital Thermometer
Each original probe, including the calibration
probe, was factory calibrated before your
Evolution System was shipped. Low-Temperature
Calibration is only needed if the calibration
values are deleted, or a new probe is installed.
Otherwise, proceed to High-Temperature
Calibration.
The following steps describe the procedure for calibration:
1. Install 7 vessels with 900 mls of water.
2. Install all Paddle blades.
3. Connect TempChek to RS-232 port (located on the side of unit) using Distek-
supplied cable.
4. Place TempChek probe in vessel to be calibrated.
5. Set vessel number on the screen using the + or button (See Figure 3-15).
6. Wait for both TempChek and vessel temperatures to stabilize.
7. Press START to calibrate.
8. Move the TempChek probe to next vessel to be calibrated.
9. Repeat step 2 through 6 until all vessels have been calibrated.
Figure 3-15: Calibrate using TempChek
Calibrate Using NIST-Traceable Reference Thermometer
Each original probe, including the calibration
probe, was factory calibrated before your
Evolution System was shipped. Low-Temperature
Calibration is only needed if the calibration
values are deleted, or a new probe is installed.
Otherwise, proceed to High-Temperature
Calibration.

Table of Contents

Related product manuals