Chapter 3: Operation
Operation Manual: Evolution Series
3
-
38
Deleting Methods from the produKEY
1. Insert the produKEY into the slot to the right of the control keypad. Turn
the key clockwise 1/4 turn. The unit will produce an audible beep.
2. The screen shown in Figure 3-32 will be displayed.
3. Highlight the method from the list and select ESC/DELETE from the keypad.
Printing a Method from the produKEY
1. Insert the produKEY into the slot to the right of the control keypad. Turn
the key clockwise 1/4 turn. The unit will produce an audible beep.
2. The screen shown in Figure 3-32 will be displayed.
3. Highlight the method from the list and select ENTER from the keypad.
4. Highlight the PRINT menu bar item and select ENTER.
Using a Premiere 5100 produKEY
It is only possible to read from a Premiere 5100
produKEY (a method cannot be written to the key).
The Evolution is capable of utilizing existing produKEYs that were previously used
on the Premiere 5100 Dissolution System.
1. Insert the produKEY into the slot to the right of the control keypad. Turn
the key clockwise 1/4 turn. The unit will produce an audible beep.
2. The screen shown in Figure 3-32 will be displayed.
3. Follow the above procedures for saving, opening, running and printing
methods.
Using the calKEY
The Evolution Dissolution System comes with a calKEY that is configured at Distek's
factory with the unit's original calibration data. In the case of a failure where the
calibration data is lost, the unit can easily restore its calibration data to the
factory settings.