Chapter 3: Operation
Operation Manual: Evolution Series
3
-
32
Figure 3-27: Manual Mode Wizard – Evolution 6100
Figure 3-28: Manual Mode Wizard – Evolution 6300
Controlling Vessels Individually (Evolution 6100 Only)
This procedure is only valid for the Evolution 6100
Bathless Dissolution System.
The Evolution 6100 Bathless Dissolution System is capable of setting test
parameters individually for each vessel. The temperature, vessel volume and
stirring element can all be varied independently for each vessel (the RPM cannot
be individually set).
1. From the main screen (see Figure 3-4) select the MANUAL MODE wizard
button.
2. Select INDIV VESSELS. Figure 3-29 will be displayed.
3. Highlight the desired vessel icon and press ENTER.
4. Figure 3-30 will now be displayed.
5. Highlight the TEMP field. Using the + and keys change to the desired test
temperature.
6. Highlight the STIR field. Using the + and keys change to the desired
stirring element (PAD for paddles or BAS for baskets).
7. Highlight the VOL field. Using the + and keys change to the desired test
volume.