Chapter 5: Troubleshooting
Operation Manual: Evolution Series
5
-
15
Error Messages
The following messages may appear on the front LCD display.
Message Possible Cause Solution
HI-CAL MEDIA TEMP IS TOO
LOW. INCREASE MEDIA TEMP
ABOVE XX°C.
The vessel temperature is
too low to perform a high
temperature calibration.
Increase media
temperature.
LO-CAL MEDIA TEMP IS TOO
HIGH. DECREASE MEDIA
TEMP BELOW XX°C.
The vessel temperature is
too high to perform a low
temperature calibration.
Decrease media
temperature.
CANNOT ADD NEW USER.
USER LIMIT HAS BEEN
EXCEEDED.
Trying to add a new user
when the maximum number
of users has already been
reached.
Delete a user prior to adding
a new user.
ONLY A MANAGER CAN ADD
NEW USERS!
A person with USER
privileges is logged into
the system. A USER does
not have correct access to
add users.
Logout of system and
have a manager log in to
add a new user.
ONLY A MANAGER CAN
DELETE A USER!
A person with USER
privileges is logged into
the system. A USER does
not have correct access to
delete users.
Logout of system and
have a manager log in to
delete a user.
ONLY A MANAGER CAN
MODIFY USER ACCOUNTS!
A person with USER
privileges is logged into
the system. A USER does
not have correct access to
modify users.
Logout of system and
have a manager log in to
modify a user.
USER: USER_NAME
EXISTS. ENTER NEW
USERNAME.
Attempting to add a user
with the same username
as an existing user.
Enter a different
username.
SYSTEM DOES NOT ALLOW
SELF DELETION!
The user was trying to
delete the login profile
that they are currently
logged in as.
You cannot delete a login
if it is currently being
used. Log out and log
back in using a different
user and then delete the
user.
ONLY A MANAGER HAS
CALIBRATION PRIVELEGES.
A person with USER
privileges is logged into
the system. A USER does
not have correct access to
calibrate the system.
Logout of system and
have a manager log in to
calibrate the system.