Chapter 3: Operation
Operation Manual: Evolution Series
3
-
39
Storing Calibration Data
The following procedure describes how to save current calibration data to the
calKEY.
1. From the Main screen select SYSTEM CONFIG.
2. From the SYSTEM screen select CALIBR.
3. Enter the valid password.
4. Insert the calKEY into the slot to the right of the control keypad. Turn the
key clockwise 1/4 turn. The unit will produce an audible beep.
5. The following message will be displayed:
UPDATE CALIBRATION
DATA FROM CALKEY.
Select NO.
6. A screen displaying the various calibration methods will be displayed
(Probe, TempChek or Reference). Select the appropriate method.
7. Select the calKEY button located on the bottom of screen.
8. The following message will be displayed:
CONFIRM SAVING OF CALIBRATION
DATA TO THE CALKEY.
Select YES.
Restoring Calibration Data
In case the calibration data is corrupted or lost, the following procedure describes
how to restore the data.
1. From the Main screen select SYSTEM CONFIG.
2. From the SYSTEM screen select CALIBR.
3. Enter the valid password.
4. Insert the calKEY into the slot to the right of the control keypad. Turn the
key clockwise 1/4 turn. The unit will produce an audible beep.
5. The following message will be displayed:
UPDATING CALIBRATION
DATA FROM THE CALKEY.
Select YES.