• K Factor : (default 1.000). This value will scale the concentration reading on the display.
• Tape Cntr : (default: off) The tape counter feature is not yet included in CLPX software.
• Keypad Beep(default on) The keypad beep feature gives audible response each time a key is pressed
and my be disabled by changing this value to off.
• Latching : (default off) The latching feature is not yet included in the CLPX software.
• BKligh Secs : (default 30) This value determines the timeout to turn off the backlight to conserve
energy.
5.4.d Config Settings 2 (Pwd Settings on older versions)
The config settings 2 menu allows more options to be modified including whether the setup menu is
password protected or not. By default any operator can access the setup menu. To limit access change this
value to ON and use the SETUP or SERVICE passwords to access the setup menu.
5.4.e Battery Settings
The battery settings allow changing of the battery limits but should only be adjusted by service personnel.
The low battery warning will appear when the battery voltage reaches the ‘Batt Warn Lvl’ or below. The
CLPX will automatically shutdown with 20 second warning when the voltage level reaches the ‘Batt
WARNING : The K Factor is multiplied by the actual concentration for
the selected gas. It is not recommended to change the value to anything
other than 1.000 .