Page 5-38 Infinity Gamma Series VF4
Message: NBP Out of Range
Parameter value: ***
NBP values are
above/below the NBP
measuring range.
Check patient and
treat, if necessary.
Verify that the inflation
mode is appropriate for
the patient. Change if
necessary.
Message: NBP Low Inflation
Limit
Parameter value: ***
The patient’s systolic
pressure is above the
current inflation limit.
Change to a higher
inflation limit.
Message: NBP Check Cuff
Size
Parameter value: ***
The wrong size NBP
cuff is being used on
the patient.
Verify the cuff size and
placement are proper
for the patient.
Verify all connections.
Change the cuff if nec-
essary.
Message: NBP Check Hose
Connection
Parameter value: ***
There is no significant
increase in cuff pres-
sure during the infla-
tion cycle.
Verify all connections.
Change the hose if
necessary.
Non-Invasive Blood Pressure (NBP) (continued)

Table of Contents

Other manuals for Dräger Infinity Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Infinity Series and is the answer not in the manual?

Dräger Infinity Series Specifications

General IconGeneral
ManufacturerDräger
Device TypePatient Monitor
Weight5 kg to 10 kg (depending on the model and configuration)
ConnectivityEthernet, USB, WLAN (optional)
Parameters MonitoredECG, SpO2, NIBP, Temperature, Respiration
Alarm SystemVisual and audible alarms, customizable alarm limits
Device CategoryMedical Device
SeriesInfinity
DisplayColor TFT
Display Size12.1 inches
Power SupplyAC Power, Internal Battery
Battery LifeUp to 4 hours (depending on configuration)

Related product manuals