Maintenance
78 Instructions for use Polaris 600
Display mounts
Component Interval Task Personnel responsible
Display mount and arm
system
Before each use Check for operational
readiness, see "Checking
for operational readiness
before each use" on page
31.
Users
Display mount and arm
system
Every 12 months
Check accompanying doc-
uments: instructions for
use available.
Service personnel
Check equipment of the
medical device for com-
pleteness according to the
instructions for use.
Perform functional and
visual inspection.
Check for deformations of
the arm system and
whether parts have
become loose.
Check for paint damage
and cracks in plastic parts.
Check whether all plastic
parts are present and in
the correct position.
Display mount and arm
system
Every 2 years
Electrical safety inspection
of the display mount
according to IEC 62353.
Replace components if
necessary.
Specialized service per-
sonnel
Check full rotation and end
positions of the arm sys-
tem.
Check the horizontal and
vertical joints of the arm
system for free movement
and re-grease them if nec-
essary.
Check the height stop of
the arm system and read-
just it if necessary.