Press the corresponding key on the keyboard to select the function. For example,
press <1> to select Data Mode; press <2> to input the wavelength for analysis.
Data Mode: The following three data mode is acceptable.
%T: (percent transmittance measurement)
ABS: (Absorbance measurement)
Conc: (Concentration measurement)
WL (nm): Determine a wavelength to be used for measurement. This
parameter can be established within a range of 190 ~ 1100 nm in
increments of 0.1 nm.
Note: There are always prompts on the bottom of the screen. You can operate
according the prompts.
3.3.1.2 Sample Setup
Press <2> to select Parameter Setup from the photometry submenu, a submenu
list appears on the right side of the screen as shown in the illustration below:
11/02/02 11:35 600.0 nm 0.854 ABS
Photometry
1 Parameter Setup 1 Name : SAMP-1
2 Sample Setup 2 Path Length ( nm ): 10.0
3 Curve Setup 3 Unit Label : MG/L
4 Print Setup
5 System Setup
6 Save Parameter
0 Forward
Select Function: _
Press <1> key to input the sample name. The name cannot be more than eight
characters. Press <2> to input the Path Length value, which means: when
measurement was preformed using a cell other than 10 mm rectangular cell, the
photometric value can be converted into one measured with a 10 mm
rectangular cell.