A(WL1)
A(WL2)
Ratio =
(b) NUM WL ‘3’ (Data Mode: Conc)
The photometric values A (WL1), A (WL2) and A (WL3) at three specified
wavelengths are calculated base on the following formula.
D = (WL2 – WL3) – (WL1 – WL3)
A(WL3) - A(WL1)
A(WL3) - A(WL2)
Ratio =
(c) When NUM WL Ranges from 2 to 6 (Data Mode: ABS or T%)
When data mode is ABS (or T%), absorbance (or transmittance) at each set
wavelength is displayed.
WL Setup
Setting the wavelength. Press <3> on the keyboard to select this function
and the setting screen will change as following.
11/12/02 15:49 600.0 nm 0.854 ABS
Multipl WL
1 Parameter Setup WL Setup (nm)
2 Sample Setup WL1(nm) 200.0
3 Curve Setup WL2(nm) 300.0
4 Print Setup WL3(nm) 400.0
5 System Setup
6 Save Parameter
7 User Baseline
0 Forward
Input Value :_
(190.0 to 1100.0)
Baseline
Press <4> key on the keyboard to select appropriate baseline type. Four
options are available: System, User1, User2 and None. System baseline is
recommended for most applications.
3-49