C3A, C6A, C6A HDVideoColposcope User Manual Product Specifications
- 93 -
Appendix 1 Product Specifications
A1.1 Environmental Specifications
Working
Temperature:
+5 ºC ~ + 40 ºC ( +41 ºF ~ +104 ºF)
Relative Humidity:
25% RH ~80% RH (non-condensing)
Atmospheric Pressure:
86kPa ~ 106kPa
Transport and Storage
Temperature:
-20 ºC ~ +55 ºC (-4ºF ~ +131 ºF)
Relative Humidity:
15% RH ~ 95% RH (non-condensing)
Atmospheric Pressure:
70kPa ~ 106kPa
A1.2 Physical Specifications
Dimensions
Vertical stand: the chassis diameter Ø 480mm
Swing arm stand: the chassis diameter Ø 550mm
MT-806 trolley: about 500mm x 500mm (length x width)
Weight
Vertical stand: 8kg
Swing arm stand: 45kg(Safe Load)
MT-806 trolley: 30kg
Power Supply
Operating Voltage: 100V-240V~
Input Current: 400mA-200mA (Power Adaptor: FW7556M/12)
Input Current: 500mA-200mA(Power Adaptor:
ME20A1202F01)
Operating Frequency: 50 Hz/60 Hz
Standards Compliance
IEC 60601-1:2005+A1:2012,EN 60601-1:2006+A12013, IEC
60601-1-2:2014, EN 60601-1-2:2015
Anti-electric Shock Type
Class II equipment(Power Adaptor: FW7556M/12)
Class I equipment (Power Adaptor: ME20A1202F01)
Working System
Continuous operation equipment
Degree of Protection against
Harmful Ingress of Water
Device/Camera: ordinary equipment (sealed equipment without
liquid proof)
Foot switch: IPX8
Degree of Safety in Presence
of Flammable Gases
Equipment not suitable for use in presence of flammable gases
Disinfection/Sterilizing
Method
Refer to this user manual for details