C3A, C6A, C6A HDVideoColposcope User Manual EMC Information
- 102 -
A4.2 Electromagnetic Immunity
Guidance and manufacture’s declaration – electromagnetic immunity
TheVideoColposcopeis intended for use in the electromagnetic environment specified
below. The customer or the user oftheVideoColposcope should assure that it is used in
such an environment.
Immunity test
IEC 60601 test
level
Compliance
level
Electromagnetic
environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
8 kV contact
15 kV air
8 kV contact
15 kV air
Floors should be wood,
concrete or ceramic tile. If
floor are covered with
synthetic material, the
relative humidity should be
at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
2 kV for power
supply lines
1 kV for
input/output lines
2kV for power
supply lines
Not Applicable
Mains power quality should
be that of a typical
commercial or hospital
environment.
Surge
IEC 61000-4-5
1 kV line(s) to
line(s)
2 kV line(s) to
earth
1 kV line(s) to
line(s)
2 kV line(s) to
earth
Mains power quality should
be that of a typical
commercial or hospital
environment.
Power frequency
(50Hz/60Hz)
magnetic field
IEC61000-4-8
30A/m
30 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.