C3A, C6A, C6A HDVideoColposcope User Manual EMC Information
- 108 -
Table-Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless
communications equipment
Test
Frequenc
y (MHz)
Brand
a)
(MHz)
Service
a)
Modulation
b)
Maximu
m
Power(W)
Distanc
e (m)
IMMUNIT
Y TEST
LEVEL
(V/m)
385
380-390
TETRA 400
Pulse
modulation
b
)
18Hz
1.8
0.3
27
450
430-470
GMRS 460,
FRS 460
FM
C)
±5 kHz
deviation
1kHz sine
2
0.3
28
710
704-787
LTE Brand 13,
17
Pulse
modulation
b
)
217 Hz
0.2
0.3
9
745
780
810
800-960
GSM
800/900,TETR
A 800, iDEN
820, CDMA
850, LTE Band
5
Pulse
modulation
b
)
18 Hz
2
0.3
28
870
930
1720
1700-199
0
GSM 1800;
CDMA 1900;
GSM 1900;
DECT; LTE
Band 1, 3, 4,25;
UMTS
Pulse
modulation
b
)
217 Hz
2
0.3
28
1845
1970
2450
2400-257
0
Bluetooth,
WLAN,802.11
b/g/n, RFID
2450, LTE
Brand 7
Pulse
modulation
b
)
217 Hz
2
0.3
28
5240
5100-580
0
WLAN 802.11
a/n
Pulse
modulation
b
)
217 Hz
0.2
0.3
9
5500
5785
Note: If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the
transmitting antenna and the ME EQUIPMENT or ME SYSTEM maybe reduce to 1m. The 1 m
test distance is permitted by IEC 61000-4-3.
a) For some services, only the uplink frequencies are included.
b) The carrier shall be modulated using a 50% duty cycle square wave signal.
c) As an alternative FM modulation, 50% pulse modulation at 18 Hz may be used because while
it does not represent actual modulation, it would be worst case

Table of Contents