EasyManua.ls Logo

Edgetech 4125 Series - Figure 5-14. Motion Sensor Calibration Result

Default Icon
145 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5-11
7. Stop the device and retrieve the current level 2 (factory) calibration using the following two
commands.
stop <CR>
getLevel2CompassCalibration <CR>
8. Ensure that the points are 0, the error percentage is 100.0%, the hard iron values are all 0.0, and
that the soft iron values are an identity matrix (see FIGURE 5-14).
Figure 5-14. Motion Sensor Calibration Result
If for some reason they are not; reset SSAHRS with the following command and then cycle power.
setLevel2CompassCalibration,factory2015,0,100.0,50.0,0,0,0,
1,0,0,0,1,0,0,0,1 <CR>

Table of Contents

Related product manuals