5. Operaon
WARNING: HIGH FILAMENT TEMPERATURE
Risk of injury or damag
e to equipment. Do not use the gauge to measure the
pressure of explosive or ammable gases or mixtures. The gauge has a heated
lament which operates approximately at 100 °C. The temperature of the lament
can be higher in fault condions.
WARNING: OVER-PRESSURISE
Risk of injury or damag
e to equipment. When you measure the pressure of gases of
high molecular weight, the pressure indicated can be below the true pressure. Make
sure that the gauge is not over-pressurised when using heavy gases.
5.1 LED indicator
When the gaug
e is connected to a power supply the LED indicator will illuminate and
ash red for approximately 5 seconds while the gauge is inialising. The LED indicator
indicates the gauge status:
▪ green if the gauge is operang normally
▪ red if an error is detected. Refer to Fault nding on page 39.
Normal operaon
During normal operaon, the LED indicator pulses to provide an approximate indicaon
of the measured pressure.
Pressure band
LED paern
Colour
Change rate
High intensity (s) Low intensity (s)
Pressure ≤ 1 mbar
Green
Connuous n/a
1 mbar < Pressure £
50 mbar
0.75 0.75
Pressure > 50 mbar 1.5 1.5
The pulsing operaon (lo
w intensity/high intensity) is switched on by default but can be
disabled (refer to Pulsed LED indicator on/o on page 34). If pulsing is disabled, the LED
will be green connuously during normal operaon.
Note:
The pulsing funcon enabled/disabled is unaect
ed by power cycling the gauge.
Error indicaon
If an error occurs within the gauge, the LED indicator will illuminate red and ash. Refer
to Fault nding on page 39.
07/2021 - ©Edwards Limited
Page 23D1G0000880_A
D1G0000880_A - Operaon