Characterisc
ID resistor
Voltage output
Filament/Pressure range
[blank] (Non-linear)
M or MP
27 kW
LC
33 kW
APG200 with matched voltage output
O, G, Q
(Log-linear with pressure)
M, MP or LC
27 kW
3.5.2 Pressure output signal
Table 10
Pressure output signal
Characterisc Pressure meas-
urement output
signal
Error voltages
Voltage output
Filament Error Voltage
APG200 with standard voltage output
X (Log-linear with pressure)
LC V=1.9 - 9.1
Filament 9.5 V
Calibraon 9.6 V
M or MP
V=1.9 - 9.1 Filament 9.5 V
2.6®1.9
1.9®2.7
Calibraon 9.6 V
[blank] (Non-linear) LC, M or MP V=2.0 - 10.0 All errors 0.1 V
APG200 with matched voltage output
O (Log-linear with pressure) LC, M or MP V=2.2 - 8.5
All errors
0.5 V
G (Log-linear with pressure) LC, M or MP V=1.9 - 10.0 0.1 V
Q (Log-linear with pressure)
LC V=1.875 - 8.875
9.5 V
M or MP V=2.573 - 8.875
3.5.3 Setpoint
Table 11
Setpoint
Characterisc Setpoint
Voltage output
Filament
Adjustment
range
Hysteresis
[blank] (Log-linear with
pressure)
LC, M or MP V=1.8 - 9.2 0.5 V
O (Log-linear with pres-
sure)
LC, M or MP V=2.0 - 8.7
10% pressure
G (Log-linear with pres-
sure)
LC, M or MP V=0.8 - 10.2
Q (Log-linear with pres-
sure)
LC, M or MP V=1.7 - 9.0
07/2021 - ©Edwards Limited
Page 15D1G0000880_A
D1G0000880_A - Technical data