Figure 9 RJ45 connector with dual relays
A. Gauge B. Connector
C. Electrical supply D. Voltmeter
A. Gauge B. Connector
C. Electrical supply D. Voltmeter
4.3.3 Maximum cable length
The maximum c
able length is dependent on the conductor cross-secon and the supply
voltage used.
Table 16
Maximum cable length
Electrical supply voltage Maximum cable length (24 AWG)
15 V 100 m
48 V 1000 m
The common mode signal voltage (that is, the voltage between the signal common and
supply common) c
an be signicant with long cables due to the voltage drop in the cable.
For accurate pressure measurements, do not connect the electrical supply ground to
signal ground. Both ground connecons are linked within the gauge.