7. Fault nding
Table 20
Faults
Fault
LED not lit on page 39
Pressure reading incorrect on page 39
Gauge shows calibraon error on page 39
Gauge shows broken lament on page 40
Fault LED not lit
Cause Incorrect electrical supply voltage. Supply polarity reversed.
Remedy Check the electrical supply and connecons.
Fault Pressure reading incorrect
Cause Vacuum leak
Remedy Do the leak check of the vacuum system.
Cause Tube has dried and requir
es adjustment
Remedy Do the atmosphere and vacuum adjustments.
Cause Tube is contaminated
Remedy Replace the tube.
Cause Gauge is measuring gas with di
er
ent thermal conducvity to Nitrogen or Air.
Remedy Refer to Gas dependency on page 29 for correcon factors.
Cause Adjustment done at an incorrect pressure
Remedy Repeat the adjustment but make sure that the pressure is at atmosphere or
vacuum.
Fault Gauge shows c
alibr
aon error
Cause Adjustment done at an incorrect pressure.
Remedy Repeat the adjustment but make sure that the pressure is at atmosphere or
vacuum.
Cause W
rong type of tube is installed.
Remedy Check that the correct type of tube is installed (M, MP, or LC).
Cause New tube has been installed.
Remedy Refer to R
emote adjustment - APG200 with transistor setpoint [blank] opon on
pag
e 36.
Cause Tube has dried outside permissible limits and can not be adjus
ted.
Remedy Replace the tube.
07/2021 - ©Edwards Limited
Page 39D1G0000880_A
D1G0000880_A - Fault nding